ADA S/N 1-Tier

Evaluating the feasibility of a 1-tiered S/N ADA testing strategy in clinical trials: A streamlined approach to immunogenicity assessment

Introduction and objective

Values verify validity – A paradigm shift in immunogenicity analysis

Current immune response detection is dominated by a multi-tier approach which requires a set of interconnected screening, confirmatory, and titration assays to provide maximum risk control. While these assays are typically applied for digital positive/negative response analysis only, they also enable a more nuanced sample status evaluation utilizing the full information of measured signal intensity values. This approach has the potential for both: simplifying assay set-up by reducing the number of assays required for the multi-tier approach – and increasing the data accessible for each sample.

In the presented data, the signal intensity approach to immunogenicity analysis was applied to a set of four different studies monitoring anti-drug-antibodies against therapeutic antibodies. All measurements were carried out with in-house optimized assays in the laboratories of Chimera. A comparison of different studies with a similar platform technology and uniform laboratory environment thereby provides a neutral and broad perspective on performance of signal intensity analysis to eventually enable a reduced number of tiers in anti-drug-antibody (“ADA”) analysis.

The key questions are:

Can a single measurement of signal intensity values in ADA detection replace the conventional multi-tier approach in immunogenicity?

Do single-tier and multi-tier results complement each other – and, if not, what is the significance of their differences?

Ask the Scientists
ADA1 tier Poster 1 Bild1

Methodology

Strength in numbers – The challenges of a robust reference value

Core hypothesis: High signal = high amount of ADA present in the sample

One measurement of signal intensity could potentially replace a multi-tier sequence of assays which provide only positive “+” / negative “-“ information.

To realize a reliable ADA analysis in a single tier approach, a set of prerequisites are required:

Appropriate increased cut-points to minimize false-positive samples:

Calculate positive/negative threshold („cut-point“) with decreased „False Positive Rate“ (e.g. 1% instead of 5% „FPR“): Thereby, confirmatory tier  assay is replaced by signal strength: positive sample status is confirmed due to high-level signals.

Minimize potential systematic bias caused by: ASSAY-TO-ASSAY VARIATION

Absolute signal intensity varies from assay to assay – Ensure and confirm during assay development that normalization (“S/N” = sample signals divided by negative control samples)  is sufficient to compensate assay-to-assay variation of measured signals

Minimize potential systematic bias caused by: DRUG INTERFERENCE

The presence of drug will decrease signals – this interferes with signal strength evaluation: Ensure and confirm improved drug tolerance during assay development and validation: While in current multi-tier approach a qualitative drug tolerance is sufficient (as long as signals are still positive decrease doesn‘t matter) one tier analysis requires quantitative drug tolerance

Minimize potential systematic bias caused by: HOOK EFFECT

A prozone effect / oversaturation with ADA will decrease signals and thus also interferes with signal strength evaluation: Ensure and confirm absence of prozone effect in the relevant concentration range (ng/mL – mg/mL) during assay development and validation; if prozone effect is there, include titration controls

Chimera has established and optimized an adapted protocol for electrochemiluminescence (MSD) detection of ADA presence in a one-step bridging assay set-up. The combination of Chimera’s AnySource® strategies and the reliable MSD technology enables robust performance, high drug tolerance, and a broad dynamic detection range.

The presence of ADA was first evaluated in a validated conventional 3-tier approach. Normalized signals (S/N) were recorded in screening assay tier for comparison with screening cut-point, screening positive samples were analyzed for signal inhibition (%SI) in confirmatory tier and confirmatory positive samples were serially diluted for titer determination, respectively.

Additionally, the comparison of the numerical value of S/N  with %SI and titer enabled determination of Pearson correlation coefficient for these parameters. Positive/negative status of S/N values were also determined with an adapted 1% FPR cut-point. The amount of samples with the same positive/negative status for 1% FPR S/N analysis vs. confirmatory status was calculated in % of total screening-positive samples.

Bild2 ADA 1 Tier
Bild3 ADA 1 Tier
Ask the Scientists

Results

Unity in diversity – Performance, patterns, and analytical potential

Data analysis demonstrated moderate to strong correlations between normalized signal intensity (“S/N”)  in screening assay and %SI of the confirmation tier. The correlation between S/N and titration tier “titer” value was typically even stronger with complete absence of systematic interference by prozone effects (which would cause an unobserved high titer but low S/N pattern). This underlined the robustness and dynamic range of the applied assay protocols.

The majority of samples (approx. 60-100%, depending on the study) revealed the same ADA positive/negative classification with either testing approach. For samples with deviations between S/N and confirmatory results, a typical signal pattern was identified. The pattern was also confirmed in additional ADA studies not included in this overview: Up to 30% of screening positive samples revealed a high/S/N but low signal interference in confirmatory assay. This could be caused by a systematic issue with the S/N value determination, indicating high signals not representing high ADA levels. If this is the case, it would be emphasizing the analytical need for a confirmatory tier test. However, it could also hint at an issue with the confirmatory tier premise which depends on drug/ADA interaction in a typically drug-tolerant set-up. If the latter is the case, parallel analysis of S/N and confirmatory tier would supplement each other and additional S/N detection would minimize the risk of misinterpretation of ADA presence.

As seen at a timeline of values measured for titer, %SI and S/N in one patient of study 2, S/N values do provide a much different kind of information as %SI: While signal strength indicates a typical increase / decrease pattern for the presence of ADA over time, %SI remains high even for supposedly smaller amounts of ADA as they are still fully quenched by the added drug in confirmatory tier set-up. If the assumption that lower S/N correctly represents lower amounts of ADA is correct, the decrease of ADA over time is much more accurately documented by S/N compared to the relatively crude titer determination: The titer here also indicates a potential decrease in ADA presence but is limited in resolution by its methodology of stepwise sample dilution.

The correlation of low amount of ADA with low S/N but still high %SI would be in favor of the interpretation that high %SI is not equivalent to high amount of antibody and thus explains the observed pattern with high S/N and low %SI in multiple studies.

These observations clearly confirm the potential additional information accessible with S/N analysis for immunogenicity monitoring. It is also obvious that establishing an appropriate comprehensive methodology benefits from the foundation of an increased dataset of a broad variety of different studies.

Ask the Scientists
ADA 1 tier Tabelle result Poster
ADA 1 tier Poster REsult Graph
ADA 1 Tier result Poster Graph 2

Conclusions

There is more than one way – Confirmed correlation is not a status, but a process

Our findings indicate that S/N analysis could support a 2-tier strategy combination of confirmatory assay and S/N evaluation. This replaces the especially elaborate titer determination (requiring multiple additional dilution samples and consuming high amounts of biological matrix). High robustness against drug and prozone interference is mandatory for this approach and should be confirmed accordingly in selection of assay set-up and validation parameters.

The observed general feasibility of a 2-tiered ADA testing strategy thereby enables more efficient sample evaluation but demonstrates also the current limitations of S/N analysis, especially in relation to the confirmatory tier.

Currently, a high correlation between %SI and S/N values is not necessarily equivalent to fully identical results for “confirmatory positive” and “S/N positive” status – and vice versa. Thus, a systematic evaluation of an increased dataset is recommended for an improved understanding of the underlying drug/ADA interactions. This is now in progress as chimera continues to subject all available ADA data to the innovative S/N analysis strategy.

Until it is fully clarified which effect causes the observed systematic pattern of low %SI vs. high S/N response, confirmatory assay %SI and “one-tier” S/N should be recorded in parallel for further in-depth analysis of signal strength information.

Results do, however, vary in impact from study to study. The evaluated studies reveal both shared correlation patterns and study-specific effects. A case-to-case evaluation of immunogenicity risk assessment and applicability of the 2-tier or even a potential 1-tier strategy should be applied during method development and validation for each study.

S/N information does provide valuable additional information for immunogenicity analysis and should thus be an important part of ADA assay set-up in the future. With expanded experience in S/N analysis, this could open a reliable pathway to the aspired streamlined 1-tier immunogenicity set-up.

The inherent potential of the 1- and 2-tier strategy to expedite data delivery and reduce resource needs in clinical development by enhanced operational efficiency clearly justifies an extended in-depth testing of this innovative approach, including further validation studies and regulatory engagement to optimize implementation in clinical trials.

Dr. Mirco Schmidt chimera biotec Assay service
Ask Dr. Mirko Schmidt

Cost efficiency

Reduce your internal costs and benefit from our specialized expertise. Our bioanalytical services include a variety of assays tailored to your specific requirements. By outsourcing your projects to us, you save costs and resources. Rely on our experience in the field of bioanalytical services.

Learn more

Time saving

Accelerate your projects with our efficient processes and fast turnaround times. Our bioanalytical services include a wide range of assays tailored to your specific needs. By outsourcing your projects to us, you save valuable time. Use our expertise for your research.

Learn more

Access to state-of-the-art technology

Take advantage of our state-of-the-art laboratories and technologies without having to invest yourself. Our bioanalytical services include a variety of assays tailored to your specific requirements. By outsourcing your projects to us, you benefit from state-of-the-art technology. Rely on our experience in the field of bioanalytical services.

Learn more

Unmatched Synergy in Ultra-Sensitive Immunoassays

Harnessing the Power of Imperacer® and AnySource® Technologies for Precise Biomarker Quantification

The combination of Chimera Biotec GmbH’s Imperacer® platform and AnySource® technology offers an outstanding synergy for ultra-sensitive immunoassays. Imperacer® enables up to 1000-fold signal amplification through its DNA-antibody conjugates, while AnySource® optimizes sample dilution and minimizes background noise. Together, these technologies allow precise analyses even with minimal sample volumes and in complex matrices. This results in high sensitivity and accuracy in biomarker quantification.

Ask our Scientist Team

ADA S/N 1-Tier

Experience the power of biotechnology – learn more now!

Unlock the potential of advanced biotech.